Table of Contents

Data Analysis

Display and measurement options

Before the different aspects of data analysis are discussed, it is shown here what different data views are available and how a measurement can be started manually. For additional information, please click here.

Examples of accurate measurements

Valid PPG Signal

The photoplethysmogram signal (PPG) displays the blood volume changes through light absorption of the arterial blood. Each cardiac cycle is shown as a peak in the raw data graph. Valid PPG signals are necessary to derive the SpO2 and respiration rate.

A good PPG signal has

Good PPG Signal

Acceleration Data

The acceleration data shows the movement of the subject. For valid data collection, especially regarding the SpO2 value and the respiration rate, movements are obstructive. In the example, the range of movement is only between 0,9 - 0,93 so the subject was almost not moving.

Good Acceleration Data

Quality Index

The quality index shows the quality of the PPG signal relating to the calculation of heart rate values. It should always be above 40 to ensure valuable calculated data. If this is not the case, it needs to be checked if the sensor has slipped out a bit of the subject's ear or if the sensor size is not correctly chosen.

Good Quality Index

Perfusion Index

The perfusion index is the quality of the PPG signal relating to the calculation of SpO2 values. It should always be in the range of 0,2 - 2. If this is not the case, another sensor size in the ear canal should be tried.

Good Perfusion Index

Accurate vital signs

When the raw data is valid, as seen in the examples above, the algorithms can calculate the vital parameters. Accurate data of SpO2, respiration rate, heart rate and body core temperature are calculated and shown in the cosinuss° WebInterface.

Good vital signs

Examples of inaccurate measurements

Bluetooth Connection Loss

If the actively measuring in-ear sensor leaves the range of 10 meters to the cosinuss° Gateway or runs out of battery, a connection loss on the graph in the cosinuss° WebInterface is visible. This is shown with straight lines through the raw data graphs, as shown in the picture below. Bluetooth Connection Loss

Wrong sensor fitting

When the core body temperature of a subject falls under 35 °C, either the sensor is slipping out of the ear or the subject has the wrong sensor size. In this case, the sensor can't detect valid raw data anymore and vital parameters potentially cannot be calculated anymore. The impact can be seen immediately on the missing SpO2 and respiration rate values.

No vital signs

Very low temperature

When the core body temperature is lowering a lot, it indicates that the sensor might have slipped out of the subject's ear or is even not inside of it. Very low temperature

Movements

Lots of movements disturb the measurements. It can be seen when the acceleration data is moving in a wide range (here from 0,7 - 1,1). As a result, the sensor might not be able to extract movement related signal changes from correct data and calculation of SpO2 and respiration rate might not be possible.

Movement Raw Data Movement Vital Signs

Limited sensor functionality

A malfunctioning or a broken sensor may have one or more of the following defects:

In most cases, the sensor will notice by itself when it is malfunctioning and the Gateway will report this to the server. This can be seen in the dashboard, when the sensor is highlighted in red:

Clicking on the sensor's serial will reveal the sensor's detailed page, where you can see the found errors. An example is provided below.

PPG Errors

If the PPG signal amplitude is below 1000 (1e+3)1) or below 2502) (depending on the firmware version) and looks like white noise, then the sensor is likely to have a defect. This can happen for only one PPG color individually or both at the same time.

In the example below, the red signal's amplitude is constantly around 7e+1 (70) while the red amplitude still shows variations

IR signal OK, red signal defect

Temperature Errors

If the temperature is unrealistically high or low, then the temperature sensor might have a defect. In some occasions, the Gateway may not start recording because the sensor's temperature is too low due to a defect.

enlightened Note: Future product generations will have even more sensory systems integrated, which means that the potential sources of error will be even more diverse. All our instructions will be updated continuously and accordingly.

1)
°Two firmware version < 3.2.0
2)
°Two firmware version >= 3.2.0